Technical index:
Main parameters | Parameter range | Resolution | Maximum allowable error |
A channel sampling flow rate | 28.3L/min | 0.1L/min | better than ±2.5% |
B channel sampling flow rate | 28.3L/min | 0.1L/min | better than ±2.5% |
Spray flow | (8~10)L/min | 0.1L/min | better than ±5.0% |
Peristaltic pump flow | (0.006~3.0)mL/min | 0.001ml/min | better than ±2.5% |
Front pressure of channel A | (-20~0)kPa | 0.01kPa | better than ±2.5% |
Front pressure of channel B | (-20~0)kPa | 0.01kPa | better than ±2.5% |
Pressure before spray flowmeter | (0~300)kPa | 0.1kPa | better than ±2.5% |
Aerosol chamber negative pressure | (-90~-120)Pa | 0.1Pa | better than ±2.0% |
Operating temperature | (0~50)℃ | ||
Cabinet negative pressure | (-50~-200)Pa | ||
Data storage capacity | >100000 group | ||
HEPA filter characteristics | Filtration efficiency for particles above 0.3um ≥99.99% | ||
Aerosol generator mass median diameter | Average particle diameter (3.0 ± 0.3) μm, geometric standard deviation ≤1.5 | ||
Dual 6-stage Anderson sampler captures particle size | Ⅰ>7μm,Ⅱ(4.7~7)μm,Ⅲ(3.3~4.7)μm,Ⅳ(2.1~3.3)μm,Ⅴ(1.1~2.1)μm,Ⅵ(0.6~1.1)μm | ||
Aerosol chamber specifications | (Length 600 x diameter 85 x thickness 3) mm | ||
Total number of positive QC sampler particles | (2200±500)cfu | ||
Negative pressure cabinet ventilation flow | ≥5m3/min |