ASTM D2303 High Voltage Tracking Index Plastic Testing Equipment

Ask Latest Price
Video Channel
Brand Name :Daxian
Model Number :DX8427
Certification :ASTM D2303,IEC60587-2007
Place of Origin :China
MOQ :One set
Price :Negotiation
Payment Terms :L/C, D/A, D/P, T/T, Western Union
Supply Ability :5 sets per month
Delivery Time :5-8 work days
Packaging Details :Wooden case
Usage :High Voltage Tracking Index Tester
Power :AC 220V/50HZ
Standards :IEC60587-2007, ASTM D2303
Drop Height :30mm ~ 40mm Adjustable
Test Voltage :100~6000V Adjustable
Warranty :12 Months
more
Contact Now

Add to Cart

Find Similar Videos
View Product Description
IEC60587 High Voltage Tracking Index Tester ASTM D2303

High polymer high voltage leakage tracking, resistance to tracking:

Silicone products have good electrical insulation properties. Their dielectric loss, voltage resistance, arc resistance, corona resistance, volume resistivity and surface resistivity are among the best in insulating materials, and their electrical properties are affected by temperature and frequency. The impact is very small. Therefore, they are a kind of stable electrical insulating materials, which are widely used in the electronics and electrical industries.
In addition to excellent heat resistance, silicone also has excellent water repellency, which is a guarantee for high reliability of electrical equipment used in wet conditions. The high-voltage tracking tester produced by Dongguan Daxian Instrument is a testing equipment developed and designed for the electrical insulation performance, tracking resistance and corrosion resistance of silicone products.

Standard:

- IEC60587-2007
- ASTM D2303

ASTM D2303 High Voltage Tracking Index Plastic Testing Equipment
Test method:
The constant tracking voltage method the constant tracking voltage is the highest voltage at which 5 samples are not damaged after 6 hours
The step-by-step voltage tracking method the step-by-step tracking voltage is the highest voltage that the five samples are not damaged after being subjected to Ih.

Instructions for use:


Note: Use a new filter paper liner for each test.

First, pour the contaminated liquid into the filter paper liner to make the filter paper fully wet. Adjust the flow rate of the contaminated liquid and correct the flow rate as specified in Table 1. Observe the flow for at least 10 minutes to ensure that the contaminated liquid flows stably on the surface of the sample between the two electrodes. The contaminated liquid should flow from the shaft hole of the upper electrode and not overflow from the side or top of the filter paper.


Preparation of the sample:

Process the sample into a sheet, preferably 6mm in thickness, 50×120mm in size, drilled with electrode mounting holes, and make a reference mark line at 45mm from the lower edge of the sample, and clean the sample with distilled or deionized water , Make the surface without any dirt, if the sample surface is smooth, it should be slightly polished;
Preparation of filter paper liner:

Use the equipped knife to cut the filter paper into the shape shown in Figure 4, and pad the cut filter paper between the upper electrode and the sample. If you use a thicker filter paper, you usually need to pad 3 to 5 layers.